Commands Listed in Alphabetical Order
ACQuire? (Query Only)
Returns the f
ollowing current acquisition parameters:
Stop after
Acquisition state
Mode
Number of average s
Sampling mode
Group
Acquisition
Syntax
ACQuire?
Related Commands
ACQuire:MODe, ACQuire:NUMACq?, ACQuire:NUMAVg,
ACQuire:STOPAfter
ACQuire:MAGnivu
Sets or returns the MagniVu feature, which provides up to 32 times signal detail
for fast viewing of short events. This feature is not recommended for slow data
formats such as RS-232.
NOTE. MagniVu channel sampling is available o n MSO oscilloscopes only.
Group
Acquisition
Syntax
ACQuire:MAGnivu {<NR1>|OFF|ON}
ACQuire:MAGnivu?
Arguments
<NR1> = 0 disables the MagniVu feature; any other value turns this feature on.
ON enables the MagniVu feature.
OFF disables the MagniVu feature.
MSO4000 and DPO4000 Series Programmer Manual 2-69