Commands Listed in Alphabetical Order
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MINIm um?
Examples
MEASUREMENT:MEAS1:MINIMUM ? might return
:MEASUREMENT:MEAS1:MINIMU M 1.75E-09.
MEASUrement:MEAS<x>:SOURCE[1]
Sets or returns the source for all single source measurements and specifies
the source to measure "from" when taking a delay mea surement or phase
measurement. Measurements are specified by <x>, which ranges from 1 through 4.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOURC E[1] {CH<x>|MATH<y>|REF<x>|D <x>}
MEASUrement:MEAS<x>:SOURC E[1]?
Arg
uments
CH<x> is an input channel waveform, where x is the channel number.
MATH<y> is a math waveform, where y is 1.
REF<x> is a reference waveform, where x is the reference channel number.
D<x> is a digital waveform, where x is the digital channel number. Digital
channels are available only on MSO models.
Examples
MEASUREMENT:MEAS2:SOUR
CE1 MATH1
specifies Math 1 as the measurement
2 source.
MEASUREMENT:MEAS1:SOU
RCE1?
might return
:MEASUREMENT:MEAS1:SOURCE [1] MATH1 indicating that Math1 is the
measurement 2 source.
MEASUrement:MEAS<x>
:SOURCE2
Sets or returns the reference source to measure "to" when taking a delay
measurement or phase measurement. Measurements are specified by <x>, which
ranges from 1 throug
h4.
Tip: Source2 measurements only a pply to phase and delay measurement types,
which require both a
target (Source1) and reference (Source2) source.
2-220 MSO4000 and DPO4000 Series Programmer Manual