Tektronix MSO4000 Landscape Lighting User Manual


 
Commands Listed in Alphabetical Order
SEARCH:SEARC
H<x>:TRIGger:A:BUS:B<x>:FLEXray:CYCLEcount:VALue
Sets or returns the binary data string to be used for FLEXRAY cycle count low
value.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A :BUS:B<x>:FLEXr ay:CYCLEcount:VALue
<QString>
SEARCH:SEARCH<x>:TRIGger:A :BUS:B<x>:FLEXr ay:CYCLEcount:
VALue?
Arguments
<QString> is a quoted string that is the binary cycle count value.
Examples
SEARCH:SEARCH1:TRIGGER:A:B US:B1:FLEXRAY:C YCLECOUNT:VALUE
“110010”
sets the cycle count value to 110010.
SEARCH:SEARCH1:TRIGGER:A:B US:B1:FLEXRAY:C YCLECOUNT:VALUE?
might return
SEARCH:SEARCH1:TRIGGER:A:B US:B1:FLEXRAY:C YCLECOUNT:VALUE
"XXXXXX"
indicating the cycle count value is don’t cares.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:FLEXray:DATa:HIVALue
Sets or returns the high binary data string used for FLEXRAY trigger if trigger
condition is ID or IDANDDATA.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A :BUS:B<x>:FLEXr ay:DATa:HIVALue
<QString>
SEARCH:SEARCH<x>:TRIGger:A :BUS:B<x>:FLEXr ay:DATa:HIVALue?
Arguments
<QString> is a quoted string that is the binary high value.
Examples
SEARCH:SEARCH1:TRIGGER:A:B US:B1:FLEXRAY:D ATA:HIVALUE
“11001010”
sets the high value 11001010.
SEARCH:SEARCH1:TRIGGER:A:B US:B1:FLEXRAY:D ATA:HIVALUE? might
return
SEARCH:SEARCH1:TRIGGER:A:BUS :B1:FLEXRAY:DATA :HIVALUE
"XXXXXXXX"
indicating the high value is don’t cares.
MSO4000 and DPO4000 Series Programmer Manual 2-341