TRIGGER:A:SET
HOLD:THRESHOLD:D1 ECL
sets the threshold to ECL levels.
TRIGGER:A:SETHOLD:THRESHO LD:D1? might return
TRIGGER:A:SE
THOLD:THRESHOLD:D1 -1.3000
indicating the
threshold is set to -1.3 volts.
TRIGger:A{:TRANsition|:RISEFall}? (Query Only)
Returns transition time trigger parameters.
Group
Trigger
Syntax
TRIGger:A{:TRANsition|:RI SEFall}?
Related Commands
TRIGger:A:UPPerthreshold:CH<x> , TRIGger:A:LOWerthreshold:CH<x>
Examples
TRIGGER:A::TRANSITION? might return
:TRIGGER:A:TRANSITION:POL ARITY POSITIVE;WHEN
SLOWER;DELTATIME 8.0000E-9
TRIGger:A{:TRANsition|:RISEFall}:DELTatime
Sets or returns the delta time used in calculating the transition value for the
transition trigger.
Group
Trigger
Syntax
TRIGger:A{:TRANsition|:RI SEFall}:DELTatim e <NR3>
TRIGger:A{:TRANsition|:RI SEFall}:DELTatim e?
Arguments
<NR3> species the delta time, in seconds.
Examples
TRIGGER:A:TRANSITION:DELT ATIME 15E-6 sets the delta time of the
transition trigger to 15 µs.
TRIGGER:A:TRANSITION:DELT ATIME? might return :TRIGGER:A
:TRANSITION:DELTATIME 2.0000E-09
indicating that the delta time of the
transition trigger is set to 2 ns.