Commands Listed in Alphabetical Order
Arguments
<NR3> is the gai
n in user units-per-division.
Examples
REF4:VERTICAL:SCALE 100E-03 sets the Reference 4 scale to 100 mV per
division.
REF4:VERTICAL:SCALE? might return :REF2:VERTIC AL:SCALE
1.0000e+00
indicating that the current vertical scale setting for Reference 2 is
1 V per division.
REM (No Query Form)
Embeds a comment within programs as a means of internally documenting the
programs. The oscilloscope ignores these embedded comment lines.
Group
Miscellaneous
Syntax
REM <QS tring>
Arguments
<QString> is a string that can contain a maximum of 80 characters.
Examples
REM "This is a comment" is a comment string that the oscilloscope will
ignore.
*RST (No Query Form)
Resets the oscilloscope to the factory default settings. The *RST command does
not alter the following:
Calibration data that affect device specications
The Output Queue
The Service Request Enable Register s etting
The Power-on status clear ag setting
Alias denitions
Stored settings
The *PUD? Response
Any o f the values associated with the DATA command.
Oscilloscope password
2-322 MSO4000 and DPO4000 Series Programmer Manual