Tektronix MSO4000 Landscape Lighting User Manual


 
Commands Listed in Alphabetical Order
Group
Waveform Trans
fer
Syntax
WFMInpre:PT_Fmt {ENV|Y}
WFMInpre:PT_Fmt?
Related Commands
WFMOutpre:PT_Fmt?
Arguments
ENV spec ies that the waveform is t ransmitted in envelope mode as maximum
and minimum point pairs. Only
Y values are explicitly transmitted. Absolute
coordinates are given by:
Xn = XZEro + XINcr (n - PT_Off)
Ynmax = YZEro + YMUlt (ynmax - YOFf)
Ynmin = YZEro + YMUlt (ynmin - YOFf)
Y species a normal waveform where one ASCII or binary data point is
transmitted for each point in the waveform record. Only
Y values are explicitly
transmitted. Absolute coordinates are given by:
Xn = XZEro + XINcr (n - PT_Off)
Yn = YZEro + YMUlt (yn - YOFf)
Examples
WFMINPRE:PT_FMT ENV sets the incoming waveform data point format to
enveloped.
WFMINPRE:PT_FMT? might return :WFMINPRE:PT_FMT ENV indicating that the
waveform is transmitted as maximum and minimum point pairs.
WFMInpre:PT_Off
The set form of this command is ignored. The query form always returns a 0. This
command is listed for compatibility w ith other Tektronix oscilloscopes.
Group
Waveform Transfer
Syntax
WFMInpre:PT_Off <NR1>
WFMInpre:PT_Off?
Arguments
Arguments are ignored.
2-482 MSO4000 and DPO4000 Series Programmer Manual