Commands Listed in Alphabetical Order
Arguments
ECL specifies a p
reset ECL high l evel of –1.3V.
TTL specifies a preset TTL high level of 1.4V.
<NR3> specifies the trigger level in volts.
TRIGger:A:LOGIc? (Query Only)
Returns all of the A logic trigger parameters.
Group
Trigger
Syntax
TRIGger:A:LOGIc?
Related Commands
TRIGger:A:LOGIc:CLAss
Examples
TRIGGER:A:LOGIC? might return :TRIGGER:A:LOGIC:CLASS
SETHOLD;FUNCTION AND;THRESHOLD:CH1 20.0000E-3;CH2 0.0000;
CH3 0.0 000;CH4 0.0000;:TRIGGER:A:LOGIC:INPUT:CH1
X;CH2 X ;CH3 X;CH4 X;CLOCK:SOURCE NONE;ED GE
RISE;:TRIGGER:A:LOGIC:PAT TERN:INPUT:CH1 X;CH2 X;CH3
X;CH4 X ;:TRIGGER :A:LOGIC:PATTERN:WHEN TRUE ;WHEN:LESSLIMIT
4.0000E-9;LIMIT 4.0000E-9;:TRIGGER:A:LOGIC:PATTERN:DELTATIME
4.0000E-9
TRIGger:A:LOGIc:CLAss
Sets or returns the class of the Logic Trigger. This command is used in conjunction
with the TRIGger:A:TYPe command.
Group
Trigger
Syntax
TRIGger:A:LOGIc:CLAss {LOGIC|SETHold}
TRIGger:A:LOGIc:CLAss?
Related Commands
TRIGger:A:TYPe, TRIGger:A:PULse:CLAss
Arguments
LOGIC sets the oscilloscope to trigger on logical combinations of the channels.
When the TRIGger:A:LOGIc:INPut:CLOCk:SOUrce is NONE,
LOGIC sets the
oscilloscope to trigger when the specified logical combinations of channels 1, 2,
2-430 MSO4000 and DPO4000 Series Programmer Manual