Commands Listed in Alphabetical Order
CH<x>:AMPSVI
AVOLTs:ENAble
Sets or returns the state of the amps via volts feature for the specified channel.
This feature supports measuring current via the voltage drop across a resistor.
Group
Vertical
Syntax
CH<x>:AMPSVIAVOLTs:ENAble {<NR1>|OFF|O N}
CH<x>:AMPSVIAVOLTs:ENAble ?
Arguments
OFF sets the amps via volts function for channel <x> to off.
ON sets the amps via volts function for channel <x> to on.
<NR1> = 0 sets the amps via volts function to off. Any other value sets the
functiontoon.
Examples
CH1:AMSVIAVOLTS:ENABLE ON turns on the amps via volts feature for channel
1.
CH<x>:AMPSVIAVOLTs:FACtor
Sets or returns the amps via volts factor for the specified channel.
Group
Vertical
Syntax
CH<x>:AMPSVIAVOLTs:FACtor <NR3>
CH<x>:AMPSVIAVOLTs:FACtor ?
Arguments
<NR3> is a double-precision ASCII string that represents the amps via volts factor.
Examples
CH1:AMPSVIAVOLTS:FACTOR 15.5 sets the amps via volts factor for channel 1
to 15.5.
CH<x>:BANdwidth
Sets or returns the selectable low-pass bandwidth limit filter for channel <x>,
where x is the channel number.
Group
Vertical
2-118 MSO4000 and DPO4000 Series Programmer Manual