Commands Listed in Alphabetical Order
TRIGger:A:BU
S:B<x>:SPI:CONDition
Sets or returns the trigger condition for a SPI trigger. Applies to bus <x>, where x
is the bus number.
Conditions
This command r equires a DPO4EMBD application module.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:SPI:CON Dition
{SS|STARTofframe|MISO|MOSI |MISOMOSI}
TRIGger:A:BUS:B<x>:SPI:CON Dition?
Related Commands
BUS:B<x>:SPI:IDLETime, BUS:B<x>:SPI:FRAMING
Arguments
SS specifies the Slave Selection condition.
STARTofframe is applicable when BUS:B<x>:SPI:FRAMING is set to
IDLEtime. When the trigger condition is set to STARTofframe,theinstrument
triggers on the first SPI clock after an idle time when there are no clocks.
MISO specifies the Master-In Slave-Out condition.
MOSI specifies the Master-Out Slave-In condition.
MISOMOSI specifies the Master-In Slave-Out and Master-Out Slave-In conditions.
TRIGger:A:BUS:B<x>:SPI:DATa{:IN|:MISO}:VALue
Sets or returns the binary data string to be used for a SPI trigger if the trigger
condition is MISO or MISOMOSI. Applies to bus <x>, where x is the bus number.
Conditions
This command r equires a DPO4EMBD application module.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:SPI:DAT a{:IN|:MISO}:VA Lue <QString>
TRIGger:A:BUS:B<x>:SPI:DAT a{:IN|:MISO}:VA Lue?
Arguments
<QString> is the binary d ata string, where the number of bits is 8 times the
number of bytes specified. The only allowed characters in the string are 0, 1,
and X.
MSO4000 and DPO4000 Series Programmer Manual 2-423