Commands Listed in Alphabetical Order
Arguments
<x> species wh
ich waveform to use, where <x> = 1 is the "from" waveform, and
<x> = 2 is the "to" waveform.
FALL species the falling edge.
RISe species the rising edge.
Examples
MEASUREMENT:MEAS1:DELAY:ED GE1 RISE species that the " from" waveform
rising edge be used for the immediate delay measurement.
MEASUREMENT:MEAS1:DELAY:ED GE1? returns either RISE or FALL.
MEASUrement:MEAS<x>:MAXimum? (Query Only)
Returns the maximum value found for this measurement since the last statistical
reset. Measurements are specied by x, which ranges from 1 through 4 .
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MAXimu m?
Examples
MEASUREMENT:MEAS3:MAXIMUM? might return
:MEASUREMENT:MEAS3:MAXIMUM 4.18E-9.
MEASUrement:MEAS<x>
:MEAN? (Query Only)
Returns the mean value accumulated for this measurement since the last statistical
reset. Measurements are specied by x, which ranges from 1 through 4 .
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MEAN?
Examples
MEASUREMENT:MEAS1:MEAN? might return :MEASUREMENT:MEA S1:MEAN
514.71E-09.
MEASUrement:MEAS<x>:MINImum? (Query Only)
Returns the minimum value for t his measurement since the last statistical reset.
Measurements are specied by <x>, which ranges from 1 through 4.
MSO4000 and DPO4000 Series Programmer Manual 2-219