Commands Listed in Alphabetical Order
TRIGger:A:PU
Lse? (Query Only)
Returns the A pulse trigger parameters.
Group
Trigger
Syntax
TRIGger:A:PULse?
Related Commands
TRIGger:A:EDGE?, TRIGger:A:LOGIc?
Examples
TRIGGER:A:PULSE? might return :TRIGGER:A:PU LSE:CLASS TRANSITION
TRIGger:A:PULse:CLAss
Sets or returns the type of pulse on which to trigger.
Group
Trigger
Syntax
TRIGger:A:PULse:CLAss {RUNt|WIDth|TRAN sition}
TRIGger:A:PULse:CLAss?
Related Commands
TRIGger:A:RUNT?, TRIGger:A:PULSEWIDth?, TRIGger:A{:TRANsition|:
RISEFall}?, TRIGger:A:TYPe
Argume
nts
RUNt triggers when a pulse crosses the first preset voltage threshold but does not
cross the second preset threshold before recrossing the first.
WIDth triggers when a pulse is found that has the specified polarity and is either
inside or outside the specified time limits.
TRANsition triggers when a pulse crosses both thresholds in the same direction
as the specified polarity and the transition time between the two threshold
crossings is greater or less than the specified time delta.
Examples
TRIGGER:A:PULSE:CLASS WIDTH specifies a width pulse for the A trigger.
TRIGGER:A:PULSE:CLASS? might return :TRIGGER:A:PULSE:CLASS WIDTH
indicating that a pulse was found that is of the specified polarity and width.
2-440 MSO4000 and DPO4000 Series Programmer Manual