Tektronix MSO4000 Landscape Lighting User Manual


 
Commands Listed in Alphabetical Order
SAVe:IMAGe:L
AYout
Sets or returns the layout to use for saved screen images.
Group
Save and Recall
Syntax
SAVe:IMAGe:LAYout {LANdscape|PORTRait}
SAVe:IMAGe:LAYout?
Arguments
LANdscape species that screen images are saved in landscape format.
PORTRait species that screen images are saved in portrait format.
SAVe:SETUp (No Query Form)
Stores the state of the oscilloscope to a specied memory location. You can later
use the *RCL command to restore the oscilloscope to this saved state.
Group
Save and Recall
Syntax
SAVe:SETUp {<file p ath>|<NR1>}
Related Commands
*RCL, RECAll:SETUp, *SAV
Arguments
<file p ath> is the target location for storing the setup le. <le path> is a
quoted string that denes the le name and path. Input the le path using the
form <drive >:<dir>/<lename>. <extension> and one or <dir>s are optional.
If you do not specify them, the oscilloscope will store the le in the current
working directory. <lename> stands for a lename. (Use of wildcard characters
in lenames is not supported.) Filename extensions are not required but are highly
recommended. For setups, use the extension ".SET".
<NR1> species a location for saving the current front-panel setup. The
front-panel setup value ranges from 1 to 10. Using an out-of-range value causes
an execution error. Any s ettings that have been stored previously at this location
will be overwritten.
Examples
SAVE:SETUP 5 saves the current oscilloscope setup in memory location 5.
SAVE:SETUP "TEK00000.SET" saves the current oscilloscope setup in the le
TEK00000.SET in the current working directory.
2-326 MSO4000 and DPO4000 Series Programmer Manual