Commands Listed in Alphabetical Order
value_in_unit
s = ((curve_in_dl - YOFf_in_dl) * YMUlt) + YZEro_in_units
NOTE. For a given w aveform record, YMUlt, YOFf, and YZEro have to be a
consistent set, otherwise vertical cursor readouts and vertical measurements may
give incorrect results.
Group
Waveform Transfer
Syntax
WFMInpre:YZEro <NR3>
WFMInpre:YZEro?
Related Commands
WFMInpre:YUNit, WFMOutpre:YZEro?
Arguments
<NR3> is the offset in YUNits.
Examples
WFMINPRE:YZERO 1.5E+0 specifies that the zero reference point for the
incoming waveform is 1.5 V below the center of the data range (given that
WFMInpre:YUNit is set to V).
WFMINPRE:YZERO? might return :WFMINPRE:YZEro 7.5000E-6 indicating
that the zero reference for the incoming waveform is 7.5 µV below the center of
the data range (given that WFMInpre:YUNit is set to V).
WFMOutpre? (Query Only)
Returns waveform transmission and formatting parameters for the waveform
specified by DATa :SOUrce command. If the waveform specified by the
DATa:SOUrce command is not displayed, the oscilloscope returns only the
waveform transmission parameters (BYT_Nr, BIT_Nr, ENCdg, BN_Fmt,
BYT_Or).
Group
Waveform Transfer
Syntax
WFMOutpre?
Examples
WFMOUTPRE? ? might return the waveform formatting data as:
:WFMOUTPRE:BYT_NR 2;BIT_NR 16;EN CDG ASCII;BN_F
MT RI;B YT_OR
MSB;WFID “Ch1, DC coupling, 100.0mV/div, 4.000us/div, 10000
points, Sample mode ”;NR_PT 10000;PT_FMT Y ;XUNIT “s”;X INCR
MSO4000 and DPO4000 Series Programmer Manual 2-487