Tektronix MSO4000 Landscape Lighting User Manual


 
Commands Listed in Alphabetical Order
Examples
SELECT:REF2 ON
turns the channel 2 waveform display on, and selects reference
waveform 2.
SELECT:REF2?
might return :SELECT:REF2 1 indicating that reference
waveform 2 is being displayed.
SET? (Query Only)
Returns the commands that list the oscilloscope s ettings except for conguration
information for the calibration values, the WFMInpre? query, and the
WFMOutpre? query. This query allows you to record or "learn" the current
oscillos
cope settings. You can use these commands to return the oscilloscope to
the state it was in when you made the SET? query. The SET? query always returns
command headers, regardless of the setting of the HEADer command. This is
because the returned commands are intended to be sent back to the oscilloscope as
a command string. The VERBose command can still be used to specify whether
the returned headers should be abbreviated or full-length.
This command is identical to the *LRN? command.
Group
Miscellaneous
Syntax
SET?
Related Commands
HEADer, *LRN?, VERBose
Examples
SET? returns a long response, part of which could be as follows: :SET
:ACQUIRE:STOPAFTER RUnsTOP;STATE 1;MODE SAMP LE;NUMENV
INFINITE;NUMAVG 16;SAMPLINGMODE RT;:HEADER 1 ;:LOCK
NONE;:LANGUAGE ENGLISH;:VERBOSE 1;:ALIAS:STATE 0;:
DISPLAY:COLOR:PALETTE NORMAL;:DISPLAY:STYLE:DOTSONLY
0;:DISPLAY:PERSISTENCE 0.0000;CLOCK 1;GRATICULE
FULL;INTEnsITY:WAVEFORM 30;GRATICULE 75;BACKLIGHT
HIGH;:HARDCOPY:INKSAVER OFF;LAYOUT LANDSCAPE;PREVI EW
0; :SAV E:IMAGE:FILEFORM AT BMP;:SAVE:WAVEFORM:FILEF ORMAT
INTERNAL;:SAVE:ASSIGN:TYPE SETUP;:TRIG GER:A:MODE
AUTO;TYPE EDGE;LEVEL 20.0000E-3;LEV EL:CH1 20.00 00E-3;CH2
0.0000;CH3 0.0000;CH4 0.0000;:TRIGG ER:A:UPPERTHRESHOLD:CH1
1.4000;CH2 800.0000E-3;CH3 800.0000 E-3;CH4 800. 0000E-3;:
TRIGGER:A:LOWERTHRESHOLD:C H1 20.0000E-3;CH2
0.0000;CH3 0.0000;CH4 0.0000;:TRIGG ER:A:HOLDOFF:TIME
20.0000E-9;:TRIGGER:A:EDGE :SOURCE CH1;COUPLING
DC;SLOPE RISE;:TRIGGER:A:LOGIC:CLASS SETHOLD;FUNCT ION
AND;THRESHOLD:CH1 20.0000E-3;CH2 0.0000;CH3
MSO4000 and DPO4000 Series Programmer Manual 2-387