Commands Listed in Alphabetical Order
Table 2-37: DATa and WFMOutpre Parameter Settings (cont.)
WFMOutpre Settings
DATa:ENCdg
Setting :ENCdg :BN_Fmt :BYT_Or :BYT_NR
RPBinary BIN RP
MSB
1,2
SRIbinary
BIN RI
LSB
1,2
SRPbinary
BIN RP
LSB
1,2
Examples
DATA:ENCDG? might return :DATa:ENCDG SRPBINARY for the format of the
outgoing waveform data.
DATA:ENCDG RPBinary sets the data encoding format to be a positive integer
where the most signicant byte is transferred rst.
DATa:SOUrce
Sets or returns the location of the waveform data transferred from the oscilloscope
by the CURVe?query.
Group
Waveform Transfer
Syntax
DATa:SOUrce
{CH1|CH2|CH3|CH4|MATH|REF1 |REF2|REF3|REF4 |D0|D1|D2|D3|D4|D5|D6|D7|D8
DATa:SOUrce?
Related Commands
CURVe
Arguments
CH1–CH4 species which analog channel data will be transferred from the
oscilloscope to the controller, channels 1 through 4.
MATH species that the Math waveform data will be transferred from the
oscilloscope to the controller.
REF1–REF4 species which Reference waveform data w ill be transferred from
the oscilloscope to the controller, waveforms, 1 through 4.
D0–D15 species which digital channel data will be transferred from the
oscilloscope to the controller, channels 0 through 15. (MSO models only)
DIGital species that the Digital waveform data will be transferred from the
oscilloscope to the controller. (MSO models only)
MSO4000 and DPO4000 Series Programmer Manual 2-149