Commands Listed in Alphabetical Order
Syntax
POWer:STATIst
ics {RESET }
Arguments
RESET cle ars the measurement statistics.
POWer:STATIstics:MODE
Enables or disables the display of the measurement statistics. Performs the same
function as the MEASUrement:STATIstics:MODE command.
Conditions
This command requires a DPO4PWR application m odule.
Group
Power
Syntax
POWer:STATIstics:MODE {OFF|ALL}
POWer:STATIstics:MODE?
Arguments
ALL turns on measurement statistics display.
OFF turns all measurements statistics off.
Examples
POWER:STATISTICS:MODE OFF turns measurements statistics display off.
POWER:STATISTICS:MODE? might return :MEASUREMENT:STATISTICS:MODE
ALL
indicating that measurement statistics are turned on and all statistics are
being displayed for each measurement.
POWer:STATIstics:WEIghting
Sets the number of samples which are included for the statistics computations
for mean and the standard deviation. Performs the same function as the
MEASUrement:STATIstics:WEIghting command.
Conditions
This command requires a DPO4PWR application m odule.
Group
Power
Syntax
POWer:STATIstics:WEIghting <NR1>;Range s {L,2,1000}
POWer:STATIstics:WEIghting ?
MSO4000 and DPO4000 Series Programmer Manual 2-297