Tektronix MSO4000 Landscape Lighting User Manual


 
Commands Listed in Alphabetical Order
Arguments
EVENTS sets the
B trigger to take place following a set number of trigger
events after the A trigger occurs. The number of events is specied by
TRIGger:B:
EVENTS:COUNt.
TIMe sets the B trigger to occur a set time a fter the A trigger event. The time
period is specied by TRIGger:B:
TIMe.
Examples
TRIGGER:B:BY TIME sets the B trigger to occur at a set time after the A trigger
event.
TRIGGER:B:BY? might return :TRIGGER:B: BY EVENTS indicating that the B
trigger takes place following a set number of trigger events after the A trigger
occurs.
TRIGger:B:EDGE? (Query Only)
Returns the source, slope, and coupling for B trigger.
Group
Trigger
Syntax
TRIGger:B:EDGE?
Related Commands
TRIGger:B:EDGE:COUPling, TRIGger:B:EDGE:SLOpe, TRIGger:B:EDGE:
SOUrce
Examples
TRIGGER:B:EDGE? might return :TRIGGER:B:EDGE:SOURCE CH1; SLOPE
RISE;COUPLING DC
TRIGger:B:EDGE:COUPling
Sets or returns the type of coupling for the B trigger.
Group
Trigger
Syntax
TRIGger:B:EDGE:COUPling {DC|HFRej|LFRe j|NOISErej}
TRIGger:B:EDGE:COUPling?
Related Commands
TRIGger:B:EDGE?
2-464 MSO4000 and DPO4000 Series Programmer Manual