Commands Listed in Alphabetical Order
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:CAN:DA Ta:VALue <QString>
TRIGger:A:BUS:B<x>:CAN:DA Ta:VALue?
Arguments
<QString> is the CAN data value in binary format. The only allowed characters
in the
QString are0,1,andX.
Examples
TRIGGER:A:BUS:B1:CAN:DATA :VALUE 1011 sets the CAN data value to 1011.
TRIGger:A:BUS:B<x>:CAN:FRAMEtype
SetsorreturnstheframetypeforaCANFRAMETypetrigger.Appliestobus
<x>, where x is the bus number.
Conditions
This command requir es a DPO4AUTO application module. This command is
only valid when the TRIGger:A:BUS:B<x>:CAN:CONDition is FRAMETYPE.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:CAN:FR AMEtype
{DATA|REMote|ERRor|OVERLo ad}
TRIGger:A:BUS:B<x>:CAN:FR AMEtype?
Arguments
DATA speciesadataframetype.
REMote species a remote
frame type.
ERRor species an error frame type.
OVERLold species an overload frame type.
Examples
TRIGGER:A:BUS:B1:CAN:FRAM ETYPE DATA sets the CAN trigger frame type
to DATA.
TRIGger:A:BUS:B<x>:CAN{:IDentier|:ADDRess}:MODe
Sets or returns the CAN a ddressing m ode for bus <x>, where x is t
he bus number.
Use this command to do the following:
2-400 MSO4000 and DPO4000 Series Programmer Manual