Commands Listed in Alphabetical Order
Syntax
MEASUrement:R
EFLevel:PERCent:MID[1] <NR3>
MEASUrement:REFLevel:PERCe nt:MID[1]?
Related Commands
MEASUrement:REFLevel:METHod
Arguments
<NR3> is the mid reference level, ranging from 0 to 100%. The default mid
reference level is 50%.
Examples
MEASUREMENT:REFLEVEL:PERCE NT:MID 1 60 sets the mid reference level
to 60% of HIGH.
MEASUREMENT:REFLEVEL:PERCE NT:MID? might return
:MEASUREMENT:REFLEVEL:PERC ENT:MID 65 indicating that the
percentage mid reference level i s set to 65% of HIGH.
MEASUrement:REFLevel:PERCent:MID2
Sets or returns the percent (where 100% is equal to HIGH) that is used to
calculate the mid reference level for the second waveform specified when
MEASUrement:REFLevel:METHod is set to Percent. This command affects the
results of delay measurements.
NOTE. this comm
and affects the associated reference level parameter for all
MEASurements:IMMed and the four periodic measurements.
Group
Measurement
Syntax
MEASUrement:REFLevel:PERCe nt:MID2 <NR3>
MEASUrement:REFLevel:PERCe nt:MID2?
Related Commands
MEASUrement:REFLevel:METHod
Arguments
<NR3> is the mid reference level, ranging from 0 to 100%. The default mid
reference level is 50%.
Examples
MEASUREMENT:REFLEVEL:PERCE NT:MID2 40 sets the mid2 reference level
to 40% of HIGH.
MEASUREMENT:REFLEVEL:PERCE NT:MID2? might return
:MEASUREMENT:REFLEVEL:PERC ENT:MID2 45 indicating that the
percentage mid2 reference level is set to 45% of HIGH.
MSO4000 and DPO4000 Series Programmer Manual 2-233