Commands Listed in Alphabetical Order
Related Commands
CURVe, DATa:ST
ARt, DATa:STOP, DATa:ENCdg, WFMInpre:NR_Pt,
WFMOutpre:NR_Pt?
Arguments
INIT initializes the waveform data parameters to their factory defaults except for
DATa:STOP, which is set to the current acquisition record length.
SNAp sets DATa:STARt and DATa:STOP to match the current waveform cursor
positions.
Examples
DATA? might return :DATA:DESTINATION REF1:ENCDG RIBINARY;SOURCE
CH1;START 1;STOP 500;WIDTH 1
DATA INIT initializes the waveform data parameters to their factory defaults.
DATa:DESTination
Sets or returns the reference memory location for storing waveform data
transferred into the oscilloscope by the CURVe command.
Group
Waveform Transfer
Syntax
DATa:DESTination REF<x>
DATa:DESTination?
Related Commands
CURVe
Arguments
REF<x> is the reference location where the waveform will be stored.
Examples
DATA:DESTINATION? might return :DATA:DESTINATION REF3 indicating that
reference 3 is the currently selected reference memory location for incoming
waveform data.
DATA:DESTINATION REF1 indicates that incoming waveform
data be stored in reference 1.
DATa:ENCdg
Sets or returns the format of outgoing waveform data. This command is equivalent
to setting WFMOutpre:ENCdg, WFMOutpre:BN_Fmt,andWFMOutpre:BYT_Or.
Setting the DATa:ENGdg value causes the corresponding WFMOutpre values
to be updated and conversley.
MSO4000 and DPO4000 Series Programmer Manual 2-147