Commands Listed in Alphabetical Order
Arguments
OFF or <NR1> =0t
urns alias expansion off. If a denedaliasissentwhen
ALIas:STATE is OFF, a command error (102) is generated.
ON or <N R1> 0tu
rns alias expansion on. When a dened alias is received, the
specied command sequence is substituted for the alias and executed.
Examples
ALIAS[:STATE] OFF turns the command alias feature off.
ALIAS[:STATE]? returns 0 when the alias feature is off.
ALLEv? (Query Only)
Prompts the oscilloscope to return all events and their messages (delimited b y
commas), and removes the returned events from the Event Queue. Use the *ESR?
query to enable the events to be returned. This command is similar to repeatedly
sending *EVMsg? queries to the oscilloscope.
Group
Status and Error
Syntax
ALLEv?
Related Commands
*ESR?, EVMsg?
Examples
ALLEV? might return :ALLEV 2225,"Measurement error, No wavefor m
to meas ure; "420,"Query UNTE RMINATED;"
AUTOSet (No Query Form)
Sets the vertical, horizontal, and trigger controls of the oscill
oscope to
automatically acquire and display the selected waveform.
Group
Miscellaneous
Syntax
AUTOSet {EXECute|UNDo}
Arguments
EXECute autosets the displayed waveform.
UNDo restores the oscilloscope settings to those present prior to the autoset
execution.
MSO4000 and DPO4000 Series Programmer Manual 2-77