Tektronix DPO4000 Landscape Lighting User Manual


 
Commands Listed in Alphabetical Order
Syntax
CH<x>:SCAle <N
R3>
CH<x>:SCAle?
Related Commands
CH<x>:OFFSet, CH<x>:POSition, REF<x>:VERTical:SCAle,
MATH[1]:VERTical:SCAle
Arguments
<NR3> is the vertical channel scale in units-per-division. The value entered here
is truncated to three signicant digits.
Examples
CH4:SCALE 100E-03 sets the channel 4 scale to 100 mV per division.
CH2:SCALE? might return :CH2:SCALE 1.0000 indicating that the current scale
setting of channel 2 is 1 V per division.
CH<x>:TERmination
Sets the connected-disconnected status of a 50 resistor, which may be connected
between the specied channel’s coupled input and oscilloscope ground. The
channel is specied by <x>. There is also a corresponding query that requests
the termination parameter and translates this enumeration into one of the two
oat values.
Group
Vertical
Syntax
CH<x>:TERmination {FIFty|MEG|<NR3>}
CH<x>:TERmination?
Arguments
FIFty sets the channel <x> input resistance to 50 .
MEG sets the channel <x> input resistance to 1 M.
<NR3> species the channel <x> input resistance numerically.
Examples
CH4:TERMINATION 50.0E+0 establishes 50 impedance on channel 1.
CH2:TERMINATION? might return :CH2:TERMINATI ON 50.0E+0 indicating
that channel 2 is set to 50 impedance.
2-130 MSO4000 and DPO4000 Series Programmer Manual