Commands Listed in Alphabetical Order
SAVe:WAVEfor
m (No Query Form)
This command saves a specified waveform or all displayed waveforms (exc l uding
serial bus waveforms). Only individual analog waveforms (CH<x>, MATH and
REF<x>) can b
e saved to reference memory locations.
You can save all displayed waveforms, excluding serial bus waveforms, to a
single CSV fi
le when the
SAVE:WAVEFORM:FILEFORMAT is set to SPREADSHEET.
You can save all displayed waveforms, excluding serial bus waveforms and digital
channels (
MSO models) to consecutive ISF (internal save format) files when the
SAVE:WAVEFORM:FILEFORMAT is set to INTERNAL.
Group
Save and Recall
Syntax
SAVe:WAVEform [<wfm>,{REF<x>}] | [<wfm>, <QString>] | [ALL,
<QString>]
Related Commands
RECAll:WAVEform, SAVe:WAVEform:FILEFormat
Arguments
<wfm>, <REF<x>> saves the specified waveform to the specified reference
memory location. <wfm> can be any live analog channel (where <x> is the
channel number), the MATH1 waveform, or another reference waveform (such
as REF1).
<wfm>, <QString> saves the specified waveform to the file specified in the
quoted string argument. Any live channel (such as CH1), the MATH1 waveform,
any reference waveform, and, for MSO models, digital waveforms (such as
D0) can be saved to a file. Digital waveforms can only be saved when the
SAVE:WAVEFORM:FILEFORMAT is set to SPREADSHEET.
ALL, <Q String> saves all displayed waveforms, excluding serial bus
waveforms, to a single CSV file specified by the quoted string argument when the
SAVE:WAVEFORM:FILEFORMAT is set to SPREADSHEET, or saves al l displayed
waveforms, excluding serial bus waveforms and digital channel waveforms (MSO
models), to individual ISF (internal save format) files with a file name prefix
specified by the argument with an underscore (_) and the waveform ID (such as
CH1, REF1, MATH) appended to the file name(s).
Examples
SAVE:WAVEFORM CH1,REF1 saves the CH1 waveform in reference memory
location 1.
:SAVE:WAVEFORM:FILEFORMAT SPREADSHEET; :SAVE:WAVEF ORM ALL,
"D:/test_folder/test1_all. csv"
saves all displayed waveforms
(excluding serial bus waveforms) to
D:/test_folder/test1_all.cs v.
MSO4000 and DPO4000 Series Programmer Manual 2-327