Commands Listed in Alphabetical Order
precise values
are required (for example, when designing to published interface
specications, such as RS-232-C).
PERCent speci
es that the reference levels are calculated as a percent
relative to HIGH and LOW. The percentages are dened using the
MEASUrement:REFLevel:PERCent commands.
Examples
MEASUREMENT:REFLEVEL:METHO D ABSOLUTE species that explicit
user-dened values are used for the reference levels.
MEASUREMENT:REFLEVEL:METHO D? might return
:MEASUREMENT:REFLEVEL:METH OD PERCENT indicating that the reference
level units used are calculated as a percent relative to HIGH and LOW.
MEASUrement:REFLevel:PERCent:HIGH
Sets or returns the percent (where 100% is equal to HIGH) used to calculate the
high reference level when MEASUrement:REFLevel:METHod is set t o Percent.
This command affects the results of rise and fall measurements.
NOTE. This command affects the associated reference level parameter for all
MEASurements:IMMed and the four periodic measurements.
Group
Measurement
Syntax
MEASUrement:REFLevel:PERCe nt:HIGH <NR3>
MEASUrement:REFLevel:PERCe nt:HIGH?
Related Commands
MEASUrement:REFLevel:METHod, MEASUrement:IMMed:TYPe,
MEASUrement:MEAS<x>:TYPe
Arguments
<NR3> is the high reference level, ranging from 0 to 100%. The default high
reference level
is 90%.
Examples
MEASUREMENT:REFLEVEL:PERCE NT:HIGH 95 sets the high reference level
to 95% of HIGH.
MEASUREMENT:REFLEVEL:PERCE NT:HIGH? might return
:MEASUREMENT:REFLEVEL:PERC ENT:HIGH 90 indicating that the
percentage high reference level is set to 90% of HIGH.
MSO4000 and DPO4000 Series Programmer Manual 2-231