Tektronix DPO4000 Landscape Lighting User Manual


 
Commands Listed in Alphabetical Order
Examples
POWER:SOA:MAS
K:MAXVOLTS 200
sets the SOA mask testing voltage to 200 V.
POWER:SOA:MASK:MAXVOLTS ? might return :PO WER:SOA:MASK:MAXVOLTS
300.0000
indi
cating that the SOA mask testing maximum voltage is set to 300 V.
POWer:SOA:MASK:MAXWatts
Sets or returns the maximum power applied to SOA mask testing.
Conditions
This command requires a DPO4PWR application m odule.
Group
Power
Syntax
POWer:SOA:MASK:MAXW
atts <N R3>
POWer:SOA:MASK:MAXWatts?
Arguments
<NR3> is the maximum power applied to SOA mask testing.
Examples
POWER:SOA:MASK:MAXWATTS 1000 sets the SOA mask testing power to
1.0000E+3 W.
POWER:SOA:MASK:MAXWATTS ? might return :PO WER:SOA:MASK:MAXWATTS
750.0000
indicating that the SOA mask testing maximum power is set to 750 W.
POWer:SOA:MASK:NR_PT? (Query Only)
Returns the number of mask points dened.
Conditions
This command requires a DPO4PWR application m odule.
Group
Power
Syntax
POWer:SOA:MASK:NR_PT?
Examples
POWER:SOA:MASK:NR_PT? might return :POW ER:SOA:MASK:NR_ PT 4
indicating that the number of mask points dened is 4.
MSO4000 and DPO4000 Series Programmer Manual 2-293