Syntax
*PSC {O FF|ON|N
R1>}
*PSC?
Related Commands
DESE, *ESE, FACtory, *RST, *SRE
Arguments
OFF sets the power-on status clear ag to false.
ON sets the power-on status clear ag to true.
<NR1> = 0 sets the power-on status clear ag to false. This disables the power-on
clear allowing the oscilloscope to possibly assert SRQ after power-on; any other
value sets the power-on status clear ag to true, enabling the power-on status clear
preventing any SRQ assertion after power on.
Examples
*PSC 0 sets t he power-on status clear ag to false.
*PSC? might return 1 to indicate that the power-on status clear ag is set to true.
*PUD
Sets or returns a string of Protected User Data. This data is protected by the
PASSWord command. You can modify it only by rst entering the correct
password. This password is not nec essary t o query the data.
Group
Status and Error
Syntax
*PUD {< Block>|<QString >}
*PUD?
Related Commands
PASSWord
Arguments
<Block> is a block containing up to 300 ASCII characters.
<QString> is a string containing up to 300 ASCII characters.
Examples
*PUD #229This oscilloscope be longs to me stores the string "This
oscilloscope belongs to me" in the user protected data area.
*PUD? might return #221PROPERTY OF CO MPANY X