Tektronix DPO4000 Landscape Lighting User Manual


 
Commands Listed in Alphabetical Order
Examples
POWER:HARMONI
CS:IEC:FILTER OFF
disables ltering of IEC harmonics.
POWER:HARMONICS:IEC:FILTER ? might return
:POWER:HARMO
NICS:IEC:FILTER 1
indicating that the IEC harmonics a re
ltered.
POWer:HARMonics:IEC:FUNDamental
Sets or retu
rns the rated fundamental current for IEC harmonics. Valid values
ranges from 0 to 16 in increments of 0.1. The unit is ampere.
Conditions
This command requires a DPO4PWR application m odule.
Group
Power
Syntax
POWer:HARMonics:IEC:FUNDam ental <NR3>
POWer:HARMonics:IEC:FUNDam ental?
Arguments
<NR3> is the fundamental current in amperes.
Examples
POWER:HARMONICS:IEC:FUNDAM ENTAL 1 sets the fundamental current to
1.0000 A.
POWER:HARMONICS:IEC:FUNDAM ENTAL? might return
:POWER:HARMONICS:IEC:FUNDA MENTAL 16 indicating that the
fundamental current for IEC harmonics is set to 16.0000 A.
POWer:HARMonics:IEC:GROUPing
Sets or returns the enabled state for grouping of IEC harmonics.
Conditions
This command requires a DPO4PWR application m odule.
Group
Power
Syntax
POWer:HARMonics:IEC:GROUPi ng {OFF|ON|1|0}
POWer:HARMonics:IEC:GROUPi ng?
MSO4000 and DPO4000 Series Programmer Manual 2-253