Tektronix DPO4000 Landscape Lighting User Manual


 
Commands Listed in Alphabetical Order
Group
Bus
Syntax
BUS:B<x>:SPI:BITOrder {LSB|MSB}
BUS:B<x>:SPI:BITOrder?
Arguments
LSB sets the bit order to least signicant bit rst.
MSB sets the bit order to m ost signicant bit rst.
Examples
BUS:B1:SPI:BITORDER LSB sets the bit order to least s ignicant bit rst.
BUS:B1:SPI:BITORDER? might return BUS:B1:SPI:BITORDER MSB
indicating the SPI bit order is set to most signicant bit rst.
BUS:B<x>:SPI{:CLOCK|:SCLK}:POLARity
Sets or returns the SPI SCLK polarity for bus <x>, where x is the bus number.
Conditions
This command requires a DPO4EMBD application module.
Group
Bus
Syntax
BUS:B<x>:SPI{:CLOCK|:SCLK }:POLARity {FALL|RISe}
BUS:B<x>:SPI{:CLOCK|:SCLK }:POLARity?
Arguments
FALL species the falling edge.
RISe species the rising edge.
BUS:B<x>:SPI{:CLOCK|:SCLK}:SOUrce
Sets or returns the SPI SCLK source for bus <x>, where x is the bus number.
Conditions
This command requires a DPO4EMBD application module.
Group
Bus
Syntax
BUS:B<x>:SPI{:CLOCK|:SCLK }:SOUrce {CH1|CH2|CH3|CH4|
D0|D1|D2|D3|D4|D5|D6|D7|D 8|D9|D10|D11|D12 |D13|D14|D15}
BUS:B<x>:SPI{:CLOCK|:SCLK }:SOUrce?
2-106 MSO4000 and DPO4000 Series Programmer Manual