Tektronix DPO4000 Landscape Lighting User Manual


 
Commands Listed in Alphabetical Order
Group
Bus
Syntax
BUS:B<1-4>:AUDio:WORDSel:S OUrce {CH1|CH2|CH3|CH4
|D0|D1|D2|D3|D4|D5|D6|D7|D 8|D9|D10|D11|D1 2|D13|D14|D15}
BUS:B<1-4>:AUDio:WORDSel:S OUrce?
Arguments
CH1–CH4 species the channel to use as t he word select source.
D0-D15 species the digital channel to use as the word select source.
Examples
BUS:B1:AUDIO:WORDSEL:SOURC E CH1 sets CH1 as the word select source.
BUS:B1:AUDIO:WORDSEL:SOURC E? might return
:BUS:B1:AUDIO:WORDSEL:SOUR CE CH2 indicating that the word select source
is set to CH2.
BUS:B<x>:CAN:BITRate
Sets or returns the bit rate for CAN bus <x>, where x is the bus number.
Conditions
This command requires a DPO4AUTO application module.
Group
Bus
Syntax
BUS:B<x>:CAN:BITRate
<NR1>|{RATE10K|RATE20K|RAT E33K|RATE37K|RA TE50K|RATE62K|
RATE83K|RATE92K|RATE100K|R ATE125K|RATE250 K|RATE500K|
RATE800K|RATE1M}
BUS:B<x>:CAN:BITRate?
Arguments
<NR1> is the bit rate. The instrument supports bit rates at 10 bps intervals. You
can enter any positive integer, and the instrument will coerce the value to the
closest supported bit rate.
Returns
The query always returns the numerical bit rate value.
Examples
bus:b1:can:bitrate rate400k sets the CAN bit rate to 400K.
bus:b1:can:bitrate?
might return :BUS:B1:CAN:BITRATE RATE800K
indicating the bit rate is set to 800K.
MSO4000 and DPO4000 Series Programmer Manual 2-91