attached probe
type. Only certain VPI probes support this command. See the
probe documentation for how to set these string arguments.
Group
Vertical
Syntax
CH<x>:PRObe:COMMAND <QString>, <QString>
Arguments
<QString> are quoted strings specifying the probe command and value to set in
the probe attached to the specied channel.
Examples
CH1:PROBE:COMMAND “MODE”, “4–4V 1M
Hz”
sets a Tektronix VPI-DPG probe
to the 4-4V1MHz mode.
CH1:PROBE:COMMAND “OUTPUT”, “ON ”
turns the output of a Tektronix
VPI-DPG probe on.
CH1:PROBE:COMMAND?“MODE” migh
treturn
CH1:PROBE:COMMAND
“MODE”,“4-4V1MHZ”
.
CH<x>:PRObe:DEGAUss (No Query Form)
Starts a degauss auto-zero cycle on a TekVPI current probe attached to the input
channel specied by <x>, where x is the channel number.
Group
Vertical
Syntax
CH<x>:PRObe:DEGAUss EXECute
Arguments
EXECute initiates the degauss operation.
CH<x>:PRObe:DEGAUss:STATE? (Query Only)
Returns the state of the probe degauss for the channel specied by <x>, where is x
is the channel number.
NOTE. This command will return
PASSED
for probes that do not support degauss
operations.