Tektronix DPO4000 Landscape Lighting User Manual


 
Commands Listed in Alphabetical Order
Syntax
CH<x>:PRObe:D
EGAUss:STATE?
Returns
NEEDED indicates the probe should be degaussed before taking measurements.
RECOMMENDED indicates the measurement accuracy might be improved by
degaussing the probe.
PASSED indicates the probe is degaussed.
FAILED indicates the degauss operation failed.
RUNNING indicates the probe degauss operation is currently in progress.
CH<x>:PRObe:FORCEDRange
Sets or returns the range of a TekVPI probe attached to the channel specied by
<x>, where x is the channel number.
NOTE. This command will return
PASSED
for probes that do not support degauss
operations.
Group
Vertical
Syntax
CH<x>:PRObe:FORCEDRange <NR3>
CH<x>:PRObe:FORCEDRange?
Arguments
<NR3> species the range, which is probe specic.
Returns
This command returns 0.0 for probes that do no support forced range.
CH<x>:PRObe:GAIN
Sets or returns the gain factor for the probe a ttached to the channel specied by
<x>, where x is the channel number. The "gain" of a probe is the output divided
by the input transfer ratio. For example, a common 10x probe has a gain of 1V.
Group
Vertical
Syntax
CH<x>:PRObe:GAIN <NR3>
CH<x>:PRObe:GAIN?
MSO4000 and DPO4000 Series Programmer Manual 2-125