Tektronix DPO4000 Landscape Lighting User Manual


 
Commands Listed in Alphabetical Order
Related Commands
MEASUrement:M
EAS<x>:TYPe
Examples
MEASUREMENT:MEAS1:UNITS? might return :MEASUREMENT:MEAS1:UNIts %
indicating units for measurement 1 are set to percent.
MEASUrement:MEAS<x>:VALue? (Query Only)
Returns a calculate value for the measurement specied by <x>, which ranges
from 1 through 4.
NOTE. This is the same value as displayed on-screen. If measurement statistics
are enabled, a new value is calculated with every waveform. In addition, this
value is updated approximately every 1/3 second. If you are acquiring a long
acquisition record, the oscilloscope may take longer to update.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:VALue ?
Related Commands
MEASUrement:MEAS<x>:UNIts?, *ESR?, ALLEv?
Examples
MEASUREMENT:MEAS1:VALUE? might return :MEA SUREMENT:MEAS1:V ALue
2.8740E-06
. If the measurement has an error or warning associated with it,
then an item is added to the error queue. The error can be checked for with the
*ESR? and ALLEv? commands.
MEASUrement:METHod
Sets or returns the method used to calculate the 0% and 100% reference level.
Group
Measurement
Syntax
MEASUrement:METHod {Auto|HIStogram|MINMax}
MEASUrement:METHod?
Related Commands
MEASUrement:REFLevel:PERCent:HIGH, MEASUrement:REFLevel:PERCent:
LOW, MEASUrement:REFLevel:PERCent:MID, MEASUrement:REFLevel:
PERCent:MID2
2-226 MSO4000 and DPO4000 Series Programmer Manual