Tektronix DPO4000 Landscape Lighting User Manual


 
Commands Listed in Alphabetical Order
Examples
BUS:B1:FLEXRA
Y:SOURCE CH4
sets the FLEXRAY source to channel 4.
BUS:B1:FLEXRAY:SOURCE? might return BU S:B1:FLEXRAY:SOURCE CH1
indicating th
e FLEXRAY source is channel 1.
BUS:B<x>:I2C:ADDRess:RWINClude
Sets and returns whether the read/write bit is included in the address.
Group
Bus
Syntax
BUS:B<x>:I2C:ADDRess:RWIN Clude {<NR1>|OFF|ON}
BUS:B<x>:I2C:ADDRess:RWIN Clude?
Arguments
<NR1> = 0 does not include the read/write bit in the addess; any other v alue
includes the read/write bit in the address..
OFF does not include the read/write bit in the address.
ON includes the read/write bit in the address.
Examples
BUS:B1:I2C:ADDRESS:RWINCL UDE ON includes the read/write bit in the
address.
BUS:B1:I2C:ADDRESS:RWINCL UDE? might return
BUS:B1:I2C:ADDRESS:RWINCL UDE 0 indicating the read/write bit is not
included in the address.
BUS:B<x>:I2C{:CLOCK|:SCLK}:SOUrce
Sets or returns the I2C SCLK source for bus <x>, where x is the bus number.
Conditions
This command requires a DPO4EMBD or DPO4COMP application module.
Group
Bus
Syntax
BUS:B<x>:I2C{:CLOCK|:SCLK }:SOUrce {CH1|CH2|CH3|CH4|
D0|D1|D2|D3|D4|D5|D6|D7|D 8|D9|D10|D11|D12 |D13|D14|D15}
BUS:B<x>:I2C{:CLOCK|:SCLK }:SOUrce?
2-96 MSO4000 and DPO4000 Series Programmer Manual