Commands Listed in Alphabetical Order
Examples
BUS:B1:FLEXRA
Y:CHANNEL B
sets the FLEXRAY id format to channel B.
BUS:B1:FLEXRAY:CHANNEL? might return BUS:B1:FLEXRAY:CHANNEL A
indicating th
at FLEXRAY id format is channel A.
BUS:B<x>:FLEXray:SIGnal
Sets or returns the FLEXRAY standard.
Group
Bus
Syntax
BUS:B<x>:FLEXray:SIGnal {BDIFFBP|BM|TX RX}
BUS:B<x>:FLEXray:SIGnal?
Arguments
BDIFFBP sets the FLEXRAY standard to BDIFFBP.
BM sets the FLEXRAY standard to BM.
TXRX sets the FLEXRAY standard to TXRX.
Examples
BUS:B1:FLEXRAY:SIGNAL BM sets the FLEXRAY standard is BM.
BUS:B1:FLEXRAY:SIGNAL? might re
turn
BUS:B1:FLEXRAY:SIGNAL
BDIFFBP
indicating the FLEXRAY standard is BDIFFBP.
BUS:B<x>:FLEXray:SOUrce
Sets or returns the FLEXRAY data
source.
Group
Bus
Syntax
BUS:B<x>:FLEXray:SOUrce
{CH1|CH2|CH3|CH4|D0|D1|D2| D3|D4|D5|D6|D7| D8|D9|D10|D11|D12|D13|D14|D
1
BUS:B<x>:FLEXray:SOUrce?
Arguments
CH<x> sets the FLEXRAY source to channel x, where x is 1 to 4.
D<x> sets the FLEXRAY source to digital channel x, where x is 1 to 16.
MSO4000 and DPO4000 Series Programmer Manual 2-95