MEASUREMENT:I
MMED:DELAY:DIRECTION FORWARDS
starts searching from the
beginning of the waveform record and looks for the rst rising or falling edge.
MEASUREMENT:
IMMED:DELAY:DIRECTION?
might return
:MEASUREMENT:IMMED:DELAY: DIRECTION BACKWARDS indicating
that searching begins at the end of the waveform record and looks for the last
rising or falling edge.
MEASUrement:IMMed:DELay:EDGE<x>
Sets or returns the slope of the edge the oscilloscope uses for the delay "from" or
"to" wave
form when taking an immediate delay measurement.
Group
Measurement
Syntax
MEASUrement:IMMed:DELay:E DGE<x> {FALL|RISe}
MEASUrement:IMMed:DELay:E DGE<x>?
Related Commands
MEASUrement:IMMed:SOUrce
MEASUrement:IMMed:SOUrce2
Arguments
<x> species which waveform to use, where <x> = 1 is the "from" waveform, and
<x>=2isthe"to"waveform.
FALL species the falling edge.
RISe species the rising edge.
Examples
MEASUREMENT:IMMED:DELAY:E DGE1 RISE species that the "from" waveform
rising edge be used for the immediate delay measurement.
MEASUREMENT:IMMED:DELAY:E DGE1? returns either RISE or FALL.
MEASUrement:IMMed:SOUrce[1]
Sets or returns the source for all single source immediate measurements
and species the source to measure "from" when taking an immediate d elay
measurement or phase measurement.
NOTE. If you do not specify a numerical sufx,thesourceisassumedtobe
SOURCE 1.