Commands Listed in Alphabetical Order
CH<x>:DESKew
Sets or returns the deskew time for channel <x>, where x is the channel number.
You can adjust the deskew time to add an independent, channel-based delay time
to the delay (
set by the horizontal position control and common to all channels)
from the common trigger point to first sample taken for each cha nnel. This lets
you compensate individual channels for different delays introduced by their
individual input hook ups.
Group
Vertical
Syntax
CH<x>:DESKew <NR3>
CH<x>:DESKew?
Arguments
<NR3> is the deskew time for channel <x>, ranging from -100 ns to +100 ns
with a resolution of 1 ps.
Examples
CH4:DESKew 5.0E-9 sets the deskew time for channel 4 to 5 ns.
CH2:DESKew? might return :CH2:DE SKEW 2.0000E-09 indicating that the
deskew time for channel 2 is se t to 2 ns.
CH<x>:INVert
Sets or returns the invert function for channel <x>, where is the channel number.
When on, the invert function inverts the waveform for the specified channel.
NOTE. This command inverts the waveform for display purposes only. The
oscilloscope does not use an inverted waveform for triggers or trigger logic inputs.
Group
Vertical
Syntax
CH<x>:INVert {ON|OFF}
CH<x>:INVert?
Arguments
OFF sets the invert function for channel <x> to off.
ON sets the invert function for channel <x> to on.
2-120 MSO4000 and DPO4000 Series Programmer Manual