Operating Basics: Functional Overview
2-8 OTS9100 User Manual
622 MHz TRIG OUT
The 622 MHz TRIG OUT provide a bit-rate/16 signal that may be
used for triggering an oscilloscope to capture the 10 Gb/s Output
signal. The output has an SMA connector and requires 50 Ohms
termination for signal integrity.
Tx DATA OUT
Tx Data Output provides a DC-coupled CML output signal for
interconnection between the module cards. This output must be
connected to the Tx DATA IN connection found on the Optics
card of the OTS9100 module using the coax cable provided.
Note: When connecting to an OTS91L1 card, a DC block
must also be installed.
10 GHz CLOCK IN
The 10 GHz Clock IN is the bit-rate clock for the Tx Data Out
interface. This input must be connected to the 10 GHz Clock Out
port of the Clock Trigger card using the coax cable provided.