Tektronix 071-0855-05 Telescope User Manual


 
Reference: Remote Commands ~ Command Description
OTS9100 User Manual 3-27
:SENSe:ANALysis:SECTion:TRACe:EXPEcted:VALUe
This command sets the data byte values used by the receiver for Trace Identifier Mismatch
measurements. The data bytes are formatted as a standard IEEE-488 data block. These values may also
be set by the :SENSe:ANALysis:SECTion:TRACe:EXPEcted command. The query form returns the
current values.
Syntax :SENSe<slot>:ANALysis:SECTion:TRACe:EXPEcted:VALUe?
:SENSe<slot>:ANALysis:SECTion:TRACe:EXPEcted:VALUe <Block>
:SENSe:ANALysis:SECTion:TRACe:EXPEcted:VALUe:STRIng
This command sets the data byte values used by the receiver for Trace Identifier Mismatch
measurements. The byte values are formatted as an ASCII string. These values may also be set by the
:SENSe:ANALysis:SECTion:TRACe:EXPEcted command. The query form returns the current values.
Syntax :SENSe<slot>:ANALysis:SECTion:TRACe:EXPEcted:VALUe:STRIng?
:SENSe<slot>:ANALysis:SECTion:TRACe:EXPEcted:VALUe:STRIng <String>
:SENSe:ANALysis:SECTion:TRACe:TIM
This command enables or disables the inclusion of Trace Identifier Mismatch errors in the analysis of
received signals. The query form returns the current setting.
Syntax :SENSe<slot>:ANALysis:SECTion:TRACe:TIM?
:SENSe<slot>:ANALysis:SECTion:TRACe:TIM <Boolean>
:SENSe:DATA:AUTOscan:STRUcture
This query returns the signal structure found in the received signal.
Syntax :SENSe<slot>:DATA:AUTOscan:STRUcture?
Parameters Description
192 Bulk data at 10Gb/s
48 Bulk data at 2.5Gb/s
12 Bulk data at 622Mb/s
3 Bulk data at 155Mb/s
1 Bulk data at 51Mb/s
0 Mixed signal structure