Tektronix 071-0855-05 Telescope User Manual


 
Reference: Remote Commands ~ Command Description
3-24 OTS9100 User Manual
:SENSe:ANALysis:PATH:HPPLM:GENEric
This command enables or disables generic equipment evaluation in the detection of Path Label Mismatch
errors. If enabled, the detection of Generic Unequipped (code 01) does not cause a Path Label Mismatch
error. The query form returns the current setting.
Syntax: :SENSe<slot>:ANALysis:PATH:HPPLM:GENEric?
:SENSe<slot>:ANALysis:PATH:HPPLM:GENEric <Boolean>
:SENSe:ANALysis:PATH:HPUNEQ
This command enables or disables the inclusion of Path Label Unequipped errors in the analysis of
received signals. The query form returns the current setting.
Syntax: :SENSe<slot>:ANALysis:PATH:HPUNEQ?
:SENSe<slot>:ANALysis:PATH:HPUNEQ <Boolean>
:SENSe:ANALysis:PATH:LABEl:EXPEcted
This command sets the comparison value for Path Label Mismatch analysis. The query form returns the
current setting.
Syntax: :SENSe<slot>:ANALysis:PATH:LABEl:EXPEcted?
:SENSe<slot>:ANALysis:PATH:LABEl:EXPEcted <NR1>
:SENSe:ANALysis:PATH:TRACe:EXPEcted
This command sets the data sequence length and data byte values against which the receiver compares
the received signal for path Trace Identifier Mismatch measurements. The data bytes are formatted as a
standard IEEE-488 data block. The query form returns the current settings of these parameters.
Syntax: :SENSe<slot>:ANALysis:PATH:TRACe:EXPEcted?
:SENSe<slot>:ANALysis:PATH:TRACe:EXPEcted <Enum> ,<Block>
Parameters Description
L1B Constant 8-bit data
L16B A repeating sequence of 16 bytes, 0-filled with CRC
L16FREE A repeating sequence of 16 bytes, arbitrary format
L64B A repeating sequence of 64 bytes, space-filled with CR/LF
L64FREE A repeating sequence of 64 bytes, arbitrary format