Tektronix 071-0855-05 Telescope User Manual


 
Reference: Remote Commands ~ Command Description
OTS9100 User Manual 3-65
:SOURce:TRIGger:MODE
This command sets the condition that will produce a trigger output from the OTS-9000 transmitter. The
query form returns the current setting.
Syntax :SOURce<slot>:TRIGger:MODE?
:SOURce<slot>:TRIGger:MODE <Enum>
Parameters Description
OFF The trigger output is disabled
ANOMaly A trigger occurs when an anomaly is generated
DEFEct A trigger occurs when a defect is generated
AD A trigger occurs when an anomaly or defect is generated
:STATus:PRESet
This command initializes the Interface Status subsystem to a defined initial state.
Syntax :SYSTem:COMMunicate:GPIB:PRIMary?
:SYSTem:COMMunicate:GPIB:PRIMary <Enum> |<NR1>
Parameters Description
NONE The GPIB port is disabled
:SYSTem:COMMunicate:GPIB:PRIMary
This command sets the primary address for the GPIB remote control port. A value of NONE disables the
GPIB port. The query form returns the current setting.
Syntax :SYSTem:COMMunicate:GPIB:PRIMary?
:SYSTem:COMMunicate:GPIB:PRIMary <Enum> |<NR1>
Parameters Description
NONE The GPIB port is disabled
:SYSTem:COMMunicate:GPIB:SECOndary
This command sets the secondary address for the GPIB port. A value of NONE disables secondary
addressing. The query form returns the current setting.
Syntax :SYSTem:COMMunicate:GPIB:SECOndary?
:SYSTem:COMMunicate:GPIB:SECOndary <Enum> |<NR1>
Parameters Description
NONE Secondary GPIB addressing is disabled