Anritsu MG369XB Portable Generator User Manual


 
Selecting a
Power Level
Sweep Range
Selecting a power level sweep range consists of
choosing a start and stop level for the power level
sweep. The power level sweep range selection pro
-
cess is identical for all power level sweep
modes—CW power sweep and sweep frequency/step
power. You can select a power level sweep range as
follows:
q
Edit the current start and stop power level pa
-
rameter values
q
Enter new start and stop power level parame
-
ter values
q
Select one of the preset power level sweep
range parameters (L1-L2, L3-L4, L5-L6, L7-L8,
or L9-L0)
Editing the Current Start / Stop Power Levels
To edit the current power level sweep range, open ei
-
ther the start or stop power level parameter. In the
display above, Edit L1 [XL1] opens the start power
level parameter and Edit L2 [XL2] opens the stop
power level parameter.
Edit the open power level parameter using the cur
-
sor control keys or the rotary data knob. When you
are finished, close the open parameter by pressing
its menu edit soft-key or by making another menu
selection.
Entering New Start / Stop Power Levels
To enter a new power level sweep range start by
opening either the start or stop power level parame
-
ters (press Edit L1 [XL1] or Edit L2 [XL2] ).
Enter a new power level using the keypad and ap
-
propriate terminator soft-key. When you are fin
-
ished, close the open parameter by pressing its
menu edit soft-key or by making another menu se
-
lection.
MG369XB OM 3-59
Local (Front Panel) Operation Power Level Sweep Operation