Anritsu MG369XB Portable Generator User Manual


 
Configuring
the GPIB
The GPIB configuration menus let you perform the
following:
q
Set the GPIB address and select the GPIB line
terminator for the signal generator
q
Select the model and set the GPIB address for
the power meter that is used to create a user
level flatness correction power-offset table
q
Select scalar mode of operation with a Wiltron
Model 562 or Anritsu Model 56100A Scalar
Network Analyzer
q
Select scalar mode of operation with a
Giga-tronics Model 8003, a Hewlett Packard
Model 8757D or a Hewlett Packard Model
8757E Scalar Network Analyzer (Only avail
-
able in units with the Pulse Modulation option
installed)
To access the Configure GPIB menu from the Sys-
tem Configuration menu, press GPIB > . The Con-
figure GPIB menu (below) is displayed.
Press GPIB Address [ADD] to change the address of
the MG369XB on the bus (the CW default GPIB ad
-
dress is five). Enter a new address, between one and
30, using the cursor control keys or the data entry
keypad and the terminator key. The new GPIB ad
-
dress will appear on the display.
Press GPIB Terminator to select a carriage return
(CR) or a carriage return and line feed (CR/LF)as
the GPIB data delimiter. Consult the GPIB control
-
ler's manual to determine which data delimiter is
required.
Press More > to access the First Additional Config
-
ure GPIB menu for more selections.
MG369XB OM 3-81
Local (Front Panel) Operation System Configuration