Anritsu MG369XB Portable Generator User Manual


 
b. Press Edit F1 to open the current fre
-
quency parameter for editing.
c. Set F1 to the first test frequency for
the model being tested.
Step 2. Verify that the Frequency Counter read
-
ing meets specifications.
Step 3. Record the Frequency Counter reading on
the test record.
NOTE
The Frequency Counter reading is
typically within ±1 Hz. Differences of
a few Hertz can be caused by noise or
counter limitations. Differences of
³±10 Hz indicate a frequency synthe
-
sis problem.
Step 4. Set F1 to the next test frequency on the
test record and record the Frequency
Counter reading.
Step 5. Repeat Step 4 until all frequencies listed
on the test record have been recorded.
5-6 MG369XB OM
Frequency Synthesis Test Operation Verification