Anritsu MG369XB Portable Generator User Manual


 
First Additional Configure GPIB Menu
When you press More >, the First Additional Config
-
ure GPIB menu (below) is displayed.
This menu lets you select the model and set the
GPIB address for the power meter that is used to
create a user level flatness correction power-offset
table. (Refer to page 3-70 for a description of this
function.
Press Power Meter Address to change the address of
the power meter on the GPIB (the power meter's de-
fault GPIB address is 13). Enter a new address, be-
tween 1 and 30, using the cursor control keys or the
data entry keypad and the terminator soft-key. The
new GPIB address will appear on the display.
Press Power Meter Select to select the power meter
model being used. (Supported power meters are the
Anritsu ML2437A, ML2438A, and ML4803A and
Hewlett-Packard 437B, 438A, and 70100A.)
Press 68/69 ID to enable operations with a Wiltron
Model 562 or Anritsu Model 56100A Scalar Network
Analyzer. (Refer to page 7-4 for master-slave proce
-
dures.) Press 68/69 ID again to disable the opera
-
tion.
Press Emulation > to access the Second Additional
Configure GPIB menu for more scalar mode of oper
-
ation choices (described on the following page).
Press Display Updates to have the display updated
with the current instrument settings when in the
remote operation mode.
Press < Previous to return to the main Configure
GPIB menu display.
3-82 MG369XB OM
System Configuration Local (Front Panel) Operation
NOTE
The Emulation > selection only ap
-
pears on this menu display in units
with the Pulse Modulation option
installed.