i. Set the sweep time to 99 seconds.
NOTE
Monitor the MG369XB’s horizontal
output on the oscilloscope to deter
-
mine sweep start and stop.
Step 4. During the analog sweep, measure the
maximum and minimum power meter
readings and record the values on the test
record. Verify that the variation (differ
-
ence between the maximum and mini
-
mum readings) does not exceed the value
noted on the test record.
Operation Verification Level Accuracy and Flatness Tests
MG369XB OM 5-15