Anritsu MG369XB Portable Generator User Manual


 
Equipment Setup
To create a power-offset table for user level flatness
calibration, connect the equipment (shown in
Figure 3-4) as follows:
Step 1. Using a GPIB cable, connect the power
meter to the MG369XB.
Step 2. Calibrate the power meter the power
sensor.
Step 3. Connect the power sensor to the point in
the test setup where the corrected power
level is desired.
Power Meter Model and GPIB Address
In order for the MG369XB to control the power me
-
ter, the GPIB address and power meter model must
be selected from the Configure GPIB menu.
Press System to access the System menu display.
At the System menu display, press Config . The Sys-
tem Configuration menu (below) is displayed.
Next, press GPIB > . The Configure GPIB menu
(below) is displayed.
MG369XB OM 3-71
Local (Front Panel) Operation Leveling Operations