Anritsu MG369XB Portable Generator User Manual


 
7-3 Use with a 56100A
Scalar Network
Analyzer
The MG369XB is directly compatible with the Anritsu Model 56100A
Scalar Network Analyzer (SNA). The following paragraphs provide in
-
structions for connecting the signal generator to the 56100A SNA so
that is can be used as a signal source for the analyzer. Operating in
-
structions for the network analyzer can be found in the Model 56100A
Scalar Network Analyzer Operation Manual, P/N 10410-00193.
Connecting
the MG369XB
to the 56100A
Connect the MG369XB signal generator to the
56100A SNA as shown in Figure 7-2.
Step 1. Connect one end of the Auxiliary I/O ca
-
ble (P/N 806-7) to the 56100A rear panel
AUX I/O connector. Connect the other end
of the cable to the MG369XB rear panel
AUX I/O connector.
Step 2. Connect one end of the dedicated system
bus cable (P/N 2100-1) to the 56100A rear
panel DEDICATED GPIB connector. Con
-
nect the other end of the cable to the
MG369XB rear panel IEEE-488 GPIB con
-
nector.
Step 3. Turn on the instrument and the 56100A.
The system is now ready to operate.
7-10 MG369XB OM
Use with a 56100A Scalar Network Analyzer Use With Other Instruments
5 6 1 0 0 A S c a l a r N e t w o r k A n a l y z e r
M G 3 6 9 X B S i g n a l G e n e r a t o r
A U X I / O
A U X I / O
I E E E - 4 8 8 G P I B
D U T
D e t e c t o r
D e d ic a t e d G P I B
S W R
A U T O T E S T E R
>
>
A
B
Figure 7-2. MG369XB to 56100A SNA Connections
NOTES
The MG369XB’s GPIB address
should be set to 5 (the default ad
-
dress setting) for operation with a
56100A SNA. To verify or change
the GPIB address setting refer to
Configuring theGPIB onpage 3-81.
The 56100A SNA will only accept
and display nine video markers, F1
thru F9, from the MG369XB.
When performing amplifier testing
only use the MG369XB power
level, L1.