Anritsu MG369XB Portable Generator User Manual


 
At the Configure GPIB menu, press More > to ac
-
cess the Additional Configure GPIB menu (below).
Press Power Mtr Address to change the address of
the power meter on the GPIB (the power meter's de
-
fault address is 13). Enter the new address, between
1 and 30, using the cursor control keys or the data
entry key pad. The new GPIB address will appear
on the display.
Press Power Mtr Select to select one of the following
power meters:
q
Anritsu ML2430A or ML4803
q
Internal (Option 8)
q
Hewlett-Packard 437, 438, or 70100A
Press < Previous to return to the main Configure
GPIB menu display.
At the Configure GPIB menu, press < Previous to
return to the System Configuration menu display.
Creating a Power-Offset Table
The MG369XB must be in CW frequency mode and
fixed (non-swept) power level mode in order to cre
-
ate a power-offset table for user level flatness cor
-
rection.
Place the signal generator in CW frequency mode by
pressing Frequency . At the resulting menu display,
press CW . The MG369XB is now in CW frequency
mode.
Place the signal generator in a fixed power level
mode by pressing Level . At the resulting menu dis
-
play, press Level . The MG369XB is now in fixed
(non-swept) power level mode.
3-72 MG369XB OM
Leveling Operations Local (Front Panel) Operation