Anritsu MG369XB Portable Generator User Manual


 
Step 3. Set up the MG369XB as follows for an an
-
alog sweep power level flatness test:
a. Reset the instrument by pressing
SYSTEM , then Reset . The CW menu
is displayed.
b.
Press Analog Sweep to place the
MG369XB in the analog sweep fre
-
quency mode and display the Analog
Sweep menu (below).
c. With the Analog Sweep menu dis-
played, press Frequency Control > .
The Sweep Frequency Control menu
(below) is displayed.
d.
Press Full to select a full range fre
-
quency sweep.
e.
Press Edit L1 to open the current
power level parameter for editing.
f. Set L1 to the power level noted on the
test record.
g. Now, return to the Analog Sweep
menu by pressing < Previous .
h. At the Analog Sweep menu, press the
menu soft-key Sweep Time to open
the sweep time parameter for editing.
5-14 MG369XB OM
Level Accuracy and Flatness Tests Operation Verification