Anritsu MG369XB Portable Generator User Manual


 
Press More > to access the Additional Step Sweep
menu (below).
This menu lets you perform the following:
q
Set the sweep time
q
Set the number of steps
q
Access the Trigger menu
q
Select log or linear sweep
q
Access the Alternate Sweep menu
Press Sweep Time [SWT] to open the sweep time
parameter.
Press Num of Steps [SNS] to open the number of
steps parameter.
Open the parameter you wish to change, then edit
the current value using the cursor control keys, the
rotary data knob, or enter a new value using the key
pad and appropriate termination soft-key. When you
have finished setting the open parameter, close it by
pressing its menu soft-key or make another menu
selection.
Press Trigger > to access the Step Sweep Trigger
menu. The trigger menu lets you select the following
sweep trigger modes:
q
Auto
q
External
q
Single
The trigger modes are described on page 3-31.
3-30 MG369XB OM
Sweep Frequency Operation Local (Front Panel) Operation
RANGE
This error message is displayed
when (1) the step size value entered
is greater than the sweep range, (2)
the number of steps entered results
in a step size of less than 0.01 Hz, or
(3) the sweep time entered divided
by the number of steps entered re
-
sults in a dwell time of <10 ms. En
-
tering valid values will clear the er
-
ror.